YIC-IR  > 科研共享资源
Potentiometric glucose sensors
Honold F ; Cammann K.
发表期刊Hormone and Metabolic Research Supplement Series
ISSN0170-5903
1988
卷号20期号:47页码:9
关键词Biosensing Techniques Catalase Electrodes Glucose/analysis** Glucose Oxidase Hydrogen-ion Concentration Oxygen/pharmacology Potentiometry
英文摘要The determination of glucose is possible with the enzymatic reaction of glucose oxidase and potentiometric detection. The signal is proportional to the concentration up to 50 mg/dl. This value is fixed by the concentration of oxygen in the sample. By adding catalase, concentrations up to 2000 mg/dl are detectable. The steepness of the calibration curve is not affected by oxygen concentrations greater than 4 mg/l. In contrast to amperometric sensors, an influence of deposits on the electrodes surface on the signal cannot be found with potentiometric sensors
文献类型期刊论文
条目标识符http://ir.yic.ac.cn/handle/133337/3009
专题科研共享资源
推荐引用方式
GB/T 7714
Honold F,Cammann K.. Potentiometric glucose sensors[J]. Hormone and Metabolic Research Supplement Series,1988,20(47):9.
APA Honold F,&Cammann K..(1988).Potentiometric glucose sensors.Hormone and Metabolic Research Supplement Series,20(47),9.
MLA Honold F,et al."Potentiometric glucose sensors".Hormone and Metabolic Research Supplement Series 20.47(1988):9.
条目包含的文件
文件名称/大小 文献类型 版本类型 开放类型 使用许可
Potentiometric gluco(172KB) 限制开放--请求全文