Improved detection limits and unbiased selectivity coefficients obtained by using ion-exchange resins in the inner reference solution of ion-selective polymeric membrane electrodes | |
Wei Qin ; Titus Zwickl ; Ernö Pretsch | |
发表期刊 | Anal. Chem. |
2000-07-15 | |
卷号 | 72期号:14页码:3236-3240 |
产权排序 | ETH, Swiss Fed Inst Technol, Dept Organ Chem, CH-8092 Zurich, Switzerland |
通讯作者 | Pretsch, E (通讯作者),ETH, Swiss Fed Inst Technol, Dept Organ Chem, Univ Str 16, CH-8092 Zurich, Switzerland |
英文摘要 | By using a high concentration of an interfering ion and a low one of the primary ion in the inner reference solution of polymeric membrane ion-selective electrodes (ISEs), the lower detection limit may be improved and unbiased thermodynamic selectivity coefficients may be obtained. To this purpose, a cation-exchange resin is used here to keep the low concentration of the primary cation constant. Different compositions of the internal solution are required for obtaining optimal lower detection limits and unbiased selectivity coefficients. All ISEs studied here, i.e., for K+, Ca2+, and NH4+, based on valinomycin, ETH 5234, and nonactin/monactin, respectively, show improved lower detection limits in the range of 10-7.6 (NH4+) to 10-8.8 M (Ca2+). Nernstian responses and, therefore, unbiased selectivity coefficients are obtained with the K+-ISE for the discriminated ions, Na+, Mg2+, and Ca2+. |
收录类别 | SCI |
语种 | 英语 |
文献类型 | 期刊论文 |
条目标识符 | http://ir.yic.ac.cn/handle/133337/5164 |
专题 | 科研共享资源 |
推荐引用方式 GB/T 7714 | Wei Qin,Titus Zwickl,Ernö Pretsch. Improved detection limits and unbiased selectivity coefficients obtained by using ion-exchange resins in the inner reference solution of ion-selective polymeric membrane electrodes[J]. Anal. Chem.,2000,72(14):3236-3240. |
APA | Wei Qin,Titus Zwickl,&Ernö Pretsch.(2000).Improved detection limits and unbiased selectivity coefficients obtained by using ion-exchange resins in the inner reference solution of ion-selective polymeric membrane electrodes.Anal. Chem.,72(14),3236-3240. |
MLA | Wei Qin,et al."Improved detection limits and unbiased selectivity coefficients obtained by using ion-exchange resins in the inner reference solution of ion-selective polymeric membrane electrodes".Anal. Chem. 72.14(2000):3236-3240. |
条目包含的文件 | ||||||
文件名称/大小 | 文献类型 | 版本类型 | 开放类型 | 使用许可 | ||
Improved detection l(123KB) | 限制开放 | 使用许可 | 请求全文 |
除非特别说明,本系统中所有内容都受版权保护,并保留所有权利。
修改评论