Potentiometric sensor based on molecularly imprinted polymer for determination of melamine in milk | |
Liang, Rongning1,2,3; Zhang, Ruiming1,3; Qin, Wei1,4 | |
发表期刊 | SENSORS AND ACTUATORS B-CHEMICAL |
2009-09-07 | |
卷号 | 141期号:2页码:544-550 |
关键词 | Molecularly Imprinted Polymer Potentiometric Sensor Melamine Ion-selective Electrode Flow Analysis |
产权排序 | Chinese Acad Sci, Environm Chem Lab, Yantai Inst Coastal Zone Res Sustainable Dev;Chinese Acad Sci, S China Sea Inst Oceanol;Chinese Acad Sci, Grad Sch;Yantai Univ, Sch Oceanol |
通讯作者 | Qin, W, Chinese Acad Sci, Environm Chem Lab, Yantai Inst Coastal Zone Res Sustainable Dev, Yantai 264003, Peoples R China |
作者部门 | 环境化学实验室 |
英文摘要 | A polymeric membrane ion-selective electrode for determination of melamine is described in this paper. it is based on a molecularly imprinted polymer (MIP) for selective recognition, which can be synthesized by using melamine as a template molecule, methacrylic acid as a functional monomer and ethylene glycol dimethacrylate as a cross-linking agent. The membrane electrode shows near-Nernstian response (54 mV/decade) to the protonated melamine over the concentration range of 5.0 x 10(-6) to 1.0 x 10(-2) mol L(-1). The electrode exhibits a short response time of similar to 16 s and can be stable for more than 2 months. Combined with flow analysis system, the potentiometric sensor has been successfully applied to the determination of melamine in milk samples. Interference from high concentrations of ions co-existing in milk samples such as K(+) and Na(+) can be effectively eliminated by on-line introduction of anion- and cation-exchanger tandem columns placed upstream, while melamine existing as neutral molecules in milk of pH 6.7 can flow through the ion-exchanger columns and be measured downstream by the proposed electrode in an acetate buffer solution of pH 3.7. (C) 2009 Elsevier BY. All rights reserved. |
文章类型 | Article |
资助机构 | Chinese Academy of Sciences [KZCX2-YW-410]; National Natural Science Foundation of China [40776058]; Ministry of Science and Technology of China [2007AA09Z103]; Department of Science and Technology of Shandong Province [2006GG2205033]; Outstanding Youth N |
收录类别 | SCI |
语种 | 英语 |
关键词[WOS] | ION-SELECTIVE ELECTRODES ; SOLID-PHASE EXTRACTION ; LIQUID-CHROMATOGRAPHY ; BIOMIMETIC SENSORS ; RECOGNITION ; MEMBRANES ; ATRAZINE ; PROTAMINE ; RESIDUES ; BINDING |
研究领域[WOS] | Chemistry ; Electrochemistry ; Instruments & Instrumentation |
WOS记录号 | WOS:000270343000030 |
引用统计 | |
文献类型 | 期刊论文 |
条目标识符 | http://ir.yic.ac.cn/handle/133337/3443 |
专题 | 中国科学院海岸带环境过程与生态修复重点实验室_海岸带环境过程实验室 中国科学院海岸带环境过程与生态修复重点实验室 |
作者单位 | 1.Chinese Acad Sci, Environm Chem Lab, Yantai Inst Coastal Zone Res Sustainable Dev, Yantai 264003, Peoples R China 2.Chinese Acad Sci, S China Sea Inst Oceanol, Guangzhou 510301, Guangdong, Peoples R China 3.Chinese Acad Sci, Grad Sch, Beijing 100049, Peoples R China 4.Yantai Univ, Sch Oceanol, Yantai 264005, Peoples R China |
推荐引用方式 GB/T 7714 | Liang, Rongning,Zhang, Ruiming,Qin, Wei. Potentiometric sensor based on molecularly imprinted polymer for determination of melamine in milk[J]. SENSORS AND ACTUATORS B-CHEMICAL,2009,141(2):544-550. |
APA | Liang, Rongning,Zhang, Ruiming,&Qin, Wei.(2009).Potentiometric sensor based on molecularly imprinted polymer for determination of melamine in milk.SENSORS AND ACTUATORS B-CHEMICAL,141(2),544-550. |
MLA | Liang, Rongning,et al."Potentiometric sensor based on molecularly imprinted polymer for determination of melamine in milk".SENSORS AND ACTUATORS B-CHEMICAL 141.2(2009):544-550. |
条目包含的文件 | 下载所有文件 | |||||
文件名称/大小 | 文献类型 | 版本类型 | 开放类型 | 使用许可 | ||
Potentiometric senso(644KB) | 开放获取 | -- | 浏览 下载 |